To remove and replace the power supplies

WA RN ING

600 watt power supply

16800 Series Portable Logic Analyzers Service Guide
Hazardous voltages exist on the power supply. To avoid electrical
shock, disconnect the power from the instrument before
performing the following procedures. After disconnecting the
power, wait at least six minutes for the capacitors on the power
supply board to discharge before servicing the instrument.
The 600 watt power supply is on all models but may look
different than the one shown here.
1
Perform previous procedures:
"To prepare the instrument for disassembly"
page 120.
"To remove and replace the cover"
Press the release latches on the sides of the connector to
2
disconnect the top 600 watt power supply cable from the
module interface board (MIB).
Press the single release tab to disconnect the lower 600
3
watt power supply cable from the MIB.
Press the release latches to disconnect the 600 watt line
4
filter cable.
Replacing Assemblies
on
on page 122.