Performance Test Record - Agilent Technologies 16800 Series Service Manual

Portable logic analyzers
Hide thumbs Also See for 16800 Series:
Table of Contents

Advertisement

3
Testing Performance

Performance Test Record

LOGIC ANALYZER MODEL NO. (circle one): 16801A, 16802A, 16803A, 16804A, 16806A, 16821A, 16822A, 16823A
Logic Analyzer Serial No.
Date:
Pulse Generator Model No.
Pulse Generator Serial No.
Pulse Generator Calibration Due Date:
Clock Rate
Pulse Generator Frequency Accuracy:
8133A: 1% of setting
Setting: 125 MHz + 2% = 127.5 MHz
250 MHz + 2% = 255 MHz
Logic Analyzer Self-Tests (Pass/Fail):
Performance Test: Minimum Master to Master Clock Time and Minimum Pulse Width
Pulse Generator Settings
Test Criteria
Pod 1 Results (pass/fail):
Pod 2 Results (pass/fail):
Pod 3 Results (pass/fail):
Pod 4 Results (pass/fail):
Pod 5 Results (pass/fail):
Pod 6 Results (pass/fail):
Pod 7 Results (pass/fail):
Pod 8 Results (pass/fail):
Pod 9 Results (pass/fail):
Pod 10 Results (pass/fail):
Pod 11 Results (pass/fail):
Pod 12 Results (pass/fail):
68
TEST EQUIPMENT USED
MEASUREMENT UNCERTAINTY
TEST RESULTS
250 Mb/s mode
Freq: 125 MHz plus test margin
Pulse Width: 1.5 ns less test margin
Test 1 of 2
eye finder locates an
Correct number of
eye for each bit
occurrences detected
Work Order No.
Recommended Test Interval - 2 Years
Recommended next testing:
Oscilloscope Model No.
Oscilloscope Serial No.
Oscilloscope Calibration Due Date:
Pulse Width (Eye Width)
Oscilloscope Horizontal Accuracy:
54845B: ±[((0.007%) (Δt)+(full scale/(2x memory
depth))+30 ps] ≅ 30 ps
Oscilloscope Display Resolution: 54845B: ±5 ps
Pulse Width setting: 1.43 ns
PWmax(worst case) = 1.43 ns+30 ps+5 ps = 1.465 ns
500 Mb/s mode (if option 500 is installed)
Freq: 250 MHz plus test margin
Pulse Width: 1.5 ns less test margin
Test 2 of 2
Test 1 of 2
eye finder locates an
eye for each bit
16800 Series Portable Logic Analyzers Service Guide
Test 2 of 2
Correct number of
occurrences detected