To remove and replace the fans

60 mm fan

Motherboard
Fan cable
16800 Series Portable Logic Analyzers Service Guide
Perform previous procedures:
1
"To prepare the instrument for disassembly"
page 120.
"To remove and replace the cover"
Disconnect the fan cable from the module interface board
2
(MIB).
3
Using a Torx T25 screwdriver, remove the 2 screws
holding the fan to the rear of the chassis.
Reverse this procedure to replace the fan.