Equipment Required - Agilent Technologies 4418B Service Manual

Power meter
Table of Contents

Advertisement

Equipment Required

Table 2-1 lists all the equipment required for the adjustments and
performance tests. If substitutions must be made, the equipment used
must meet the critical specifications.
Equipment
Required
Digital Multimeter
DC Calibrator
Range Calibrator
Frequency Counter
Power Meter
Thermistor Mount
Power Sensor
30 dB Attenuator
Power Sensor
Cable
1. P = Performance Tests, A = Adjustments
Agilent E4418B/E4419B Service Guide
Table 2-1: Required Equipment
Model Number
Agilent 3458A
Datron 4000A
or
Fluke 5700A/5720A
Agilent 11683A
or
Agilent 11683A
Opt H01
Agilent 53132A
Agilent 432A
Agilent 478A Opt H75/H76
or
Agilent 8478B Opt H01
Agilent E-series power
sensor
Agilent 11708A
or
Agilent 8491A Option 030
Agilent 11730A
or
Agilent
E9288A
Critical Specification
Range 0 to 20 V
Resolution 0.01 V
Option H01 is essential for
accurate verification. (See
analysis in
Appendix B
). The
Standard 11683A can be used
for indication only.
Range 10 Hz to 50 MHz
Resolution 1 Hz
Range 1 mW
Transfer Accuracy 0.2%
SWR<1.05 at 50 MHz
Accuracy ±0.2% at 50 MHz.
Performance Tests
Introduction
1
Usage
P and A
P
P
P and A
P and A
P and A
P
P
P and A
2-3