3
Testing Performance

Determine PASS/FAIL (2 of 2 tests)

NO TE
54
Pass/Fail Point: The Listing window is set up to search for
the appropriate number of A's and 5's in the acquisition. If
the logic analyzer does not detect the correct number of A's
and 5's, an error dialog will appear.
1
Click the Run Repetitive icon
run for about one minute. The analyzer will acquire data
and the Listing window will continuously update.
If the "can't find occurrence" dialog appears, then the
logic analyzer fails the test.
Check your test setup. If the failure is not the result of a
problem with the test setup, record the failure in the "Test
2 of 2: Correct number of occurrences detected" section of the
Performance Test Record
Be sure that the black ground lead is making good contact with the ground
pin on the test connector.
When about one minute has elapsed, click the Stop button
2
to stop the acquisition.
If the "can't find occurrence" dialog does not appear, then
the analyzer passes the test. Record "Pass" in the "Test 2 of
2: Correct number of occurrences detected" section of the
Performance Test Record
. Let the logic analyzer
(page
68).
(page
68).