Troubleshooting
Configure the logic analyzer to test Pod 1
106
Exit the logic analysis application (from the main menu,
1
choose File→Exit) and then restart the application. This
puts the logic analyzer into its initial state.
Set up the bus and signals to test Pod 1.
2
From the Agilent Logic Analyzer application's main
a
menu, select Setup→Bus/Signal... or Setup→My Logic
Analyzer-1→Bus/Signal....
In the Analyzer Setup dialog, ensure that the Threshold
b
button for Pod 1 is set to TTL (1.50 V).
Threshold
Verify that the activity indicators (the red arrows)
c
show activity on all 16 channels that are connected to
the pulse generator.
d
Assign all channels. Hint: you can do this quickly by
clicking on the left- most channel, then dragging to the
right across all of the other channels. If you have a
touch screen model (Option 103), you can touch the
touchscreen and drag across with your finger.