Range
32.999 mV
32.999 mV
32.999 mV
32.999 mV
32.999 mV
32.999 mV
32.999 mV
32.999 mV
32.999 mV
32.999 mV
32.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
30
Table 7. Verification Tests for AC Voltage (Normal)
Output
Frequency
3.000 mV
45 Hz
3.000 mV
10 kHz
30.000 mV
9.5 Hz
30.000 mV
10 Hz
30.000 mV
45 Hz
30.000 mV
1 kHz
30.000 mV
10 kHz
30.000 mV
20 kHz
30.000 mV
50 kHz
30.000 mV
100 kHz
30.000 mV
450 kHz
33.000 mV
45 Hz
33.000 mV
10 kHz
300.000 mV
9.5 Hz
300.000 mV
10 Hz
300.000 mV
45 Hz
300.000 mV
1 kHz
300.000 mV
10 kHz
300.000 mV
20 kHz
300.000 mV
50 kHz
300.000 mV
100 kHz
300.000 mV
500 kHz
0.33000 V
45 Hz
0.33000 V
10 kHz
3.00000 V
9.5 Hz
3.00000 V
10 Hz
3.00000 V
45 Hz
3.00000 V
1 kHz
3.00000 V
10 kHz
3.00000 V
20 kHz
3.00000 V
50 kHz
Lower Limit
Upper Limit
2.977 mV
3.022 mV
2.977 mV
3.022 mV
28.350 mV
31.650 mV,
29.944 mV
30.056 mV
29.956 mV
30.044 mV
29.956 mV
30.044 mV
29.956 mV
30.044 mV
29.944 mV
30.056 mV
29.932 mV
30.068 mV
29.877 mV
30.123 mV
29.715 mV
30.285 mV
32,970 mV
33.029 mV
32.970 mV
33.029 mV
283.350 mV
316.650 mV
299.917 mV
300.083 mV
299.893 mV
300.107 mV
299.983 mV
300.107 mV
299.983 mV
300.107 mV
299.782 mV
300.218 mV
299.702 mV
300.298 mV
299.311 mV
300.689 mV
298.470 mV
301.530 mV
0.32984 V
0.33015 V
0.32984 V
0.33015V
2.83500 V
3.16500 V
2.99868 V
3.00132 V
2.99910 V
3.00090 V
2.99910V
3.00090 V
2.99910 V
3.00090 V
2.99817 V
3.00183 V
2.99745 V
3.00255 V