Dc Voltage Output - Fluke 5730A Operator's Manual

Multifunction calibrator
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A
Operators Manual

DC Voltage Output

To set a dc voltage output:
1. Make sure the Calibrator is in standby (STANDBY annunciator lit). Push S if
2. If the UUT is not connected, connect it now as described in this chapter under
3. Set the UUT to measure dc voltage on the necessary range.
4. Enter a voltage value with the numeric keypad.
5. To change the polarity of the entry, push +.
6. Push U, N, or K if necessary.
7. Push V.
8. The display now shows the amplitude of the entry. If an entry error is made, push C
9. Push E. The Calibrator clears the entry from the entry bar toward the bottom
10. Push Pto activate the Calibrator output. The UUT will respond to the applied
4-16
necessary.
"Connect the Calibrator to a UUT."
to clear the display, and then reenter the value. If the most recent digit entered is in
error, push B to clear that digit. The display shown below assumes an entry of
10 mV.
and shows in the larger area above. No voltage is available at the output terminals
until P is pushed.
voltage.
In dc voltage output mode, the selections available at the bottom are:
External Sense
External Guard
Wideband
hhp113.tif