Leakage Current Source; Rcd (Residual Current Device); Ac/Dc Voltage Calibrator (5320A/Vlc Only) - Fluke 5320A Getting Started Manual

Multifunction electrical tester calibrator
Hide thumbs Also See for 5320A:
Table of Contents

Advertisement

5320A
Getting Started Manual
Correction COMP Mode (Active Loop Compensation) (5320A/VLC only)
Residual Impedance Range ............................. 0 to 2 Ω
Maximum Test Current ..................................... <25/N A pk, where N equals number of UUT generated test current
Uncertainty of compensation .......................... (1 % + 15 mΩ) + uncertainty of selected resistance value. Uncertainty

Leakage Current Source

Range ................................................................. 0.1 to 30 mA
Resolution:
Passive Mode ................................................. 10 μA setting, 1 μA measurement
Differential Mode............................................. 10 μA setting, 1 μA measurement
Substitute Mode.............................................. 10 μA
Active Mode (5320A/VLC only)....................... 10 μA
Test Voltage:
Passive Mode ................................................. 60 to 250 V ac + dc rms
Differential Mode............................................. 60 to 250 V ac + dc rms
Substitute Mode.............................................. 10 to 250 V ac + dc rms
Active Mode (5320A/VLC only)....................... 50 to 100 V ac + dc rms
Uncertainty:
Passive Mode ................................................. 0.3 % + 2 μA ac + dc rms
Differential Mode............................................. 0.3 % + 2 μA ac + dc rms
Substitute Mode.............................................. 0.3 % + 2 μA ac + dc rms
Active Mode (5320A/VLC only)....................... 0.3 % + 1 μA ac + dc rms

RCD (Residual Current Device)

Trip Current Range:
0.5 X I and 1 X I Mode .................................... 3 to 3000 mA in 1 mA steps
1.4 X I and 2 X I Mode .................................... 3 to 1500 mA in 1 mA steps
5 X I Mode ...................................................... 3 to 600 mA in 1 mA steps
Trip Current Measurement Resolution ........... 1 μA on 30 mA range
Uncertainty:
0.5 X I and 1 X I Mode .................................... 1 % rms
1.4 X I and 2 X I Mode .................................... 2 % rms
5 X I Mode ...................................................... 5 % rms
Trip Time Range................................................ 10 to 5000 ms
Trip Time Uncertainty....................................... 0.02 % + 0.25 ms
Series Resistance ............................................. 0.025 Ω, 0.05 Ω, 0.1 Ω, 0.33 Ω, 0.5 Ω, 1 Ω, 1.8 Ω, 5 Ω, 10 Ω, 18 Ω,
Line/Touch Voltage Range............................... 250 V
Line/Touch Voltage Uncertainty...................... 5 % + 3 V

AC/DC Voltage Calibrator (5320A/VLC only)

Range ................................................................. 3 to 600 V, ac or dc
Resolution ......................................................... 4 digits
Internal Ranges:
AC Mode......................................................... 30, 100, 300, and 600 V (Autoranging only)
DC Mode......................................................... 30, 150, and 600 V (Autoranging only)
Frequency:
Range ............................................................. 40 to 400 Hz
Resolution....................................................... 3 digits
8
periods.
is valid at the point in time when the COMP function is initiated.
Test uncertainty can be influenced by power line voltage instability
10 μA on 300 mA range
100 μA on 3A range
50 Ω, 100 Ω, 180 Ω, 500 Ω, 1000 Ω, 1800 Ω