2285646 - Through Bolts Without Insert - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
2.3.3.2 2285646 – Through bolts without insert
Table 5-9:
Item
Name
Bolt, Hex
Washer, Flat
Plate
Nut, Hex
Grout assy
Dowel
Chapter 5 Physical Characteristics
Part #
2296892
Through Bolt; M20 - 500-400
12 bolts for LC Positioner and 4 bolts
for Table
99142204
Special Flat Washer for Through
Bolts; one for each bolt
2290941
Special Steel Spacer Plate; 4 in. x
4 in. (102 mm x 102 mm); one for
each bolt
99141607
Hex Nut M20 STL galvanized two
for each bolt
2285055
- 10 kg Powdered Mortar Ardex K-15
- RTV Silicon Rubber Adhesive
- 18 mm making tape adhesive
- Open cell foam
2290937
Wood Dowel; 24 mm diameter
Direction 5314528-2-1EN, Revision 1
Description
16 pc
16 pc
16 pc
32 pc
1 kit
12 pc
Quan.
Notes
Through Bolt mounting
method hardware
same as above
same as above
same as above
Used in constructing LC
Positioner grout dam
141

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3100Innova 3100-iqInnova 4100Innova 4100-iq

Table of Contents