Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
10 Emergency
10.1 Partial UPS Fluoro (option)
A Fluoro UPS (20 kVA) has been designed for the Innova system. This partial UPS lets
the customer complete an exam in fluoro mode in case of a power failure. The autonomy
provided by this UPS is 5 minutes of fluoro every 24 hours.
10.2 Full UPS Record (option)
An UPS sized for 150 kVA is required to supply the system in record mode. Such an UPS
will provide to the customer about 10 minutes of autonomy. Contact your local GEMS
representative for more information and/or public power distribution analysis.
An UPS is inserted inside C1 Cabinet and the DL is not affected by microcuts with a duration
NOTE:
of less than 10µs.
C1 UPS accepts a power cut without a hardware risk.
The UPS inserted inside C1 Cabinet is a part of the system and it is not permitted to replace
it by another locally supplied unit.
UPS autonomy is 1 minute before shutdown. Time before shut down is 1 minute without risk
of loosing stored image.
Time to completely charge the battery of UPS is 48 hours.