Route separately EMI filter incoming power lines and outgoing power lines (refer to
Electrical Connections).
NOTE:
The maximum distance between the EMI filter and the PDB is 3 m.
NOTE:
The Fluoro UPS (CE) option requires an EMI filter box placed upstream the PDB. The EMI
filter box will be delivered as a single product by GE Consumer & Industrial. For further
information on the EMI filter box , see Typical EMI Filter box in

2.5 Radiation Protection

Because x–ray equipment produces radiation, special precautions may be needed or
special site modifications may be required. The General Electric Company does not make
recommendations regarding radiation protection. It is the purchasers' responsibility to consult
a radiation physicist for advise on radiation protection in x–ray rooms.

2.6 Audible noise

Less than 50 dB (A) at 1 meter for an Innova LC Positioner.
Limited to 50 dB (A) at 1 meter for Omega IV, Omega V and Tilt tables.
Limited to 55 dB (A) at 1 meter for C2 Cabinet.
Limited to 60 dB (A) at 1 meter for the COOLIX 4000.
Limited to 65 dB (A) at 1 meter for C1 Cabinet.
Limited to 60 dB (A) at 1 meter for Digital Detector Chiller Thermo–Tek.
Limited to 52 dB (A) (background of 35 dB (A)) at 1 meter for Digital Detector
Chiller Thermo–Con.
Less than 50 dB (A) at 1 meter for a DL LCD monitor.
Less than 60 dB (A) at 1 meter for the Fluoro UPS.
NOTE:
Noise generated by system cabinets all together is 70dB. If cabinets are installed too close to
the exam room (e.g. behind sliding doors) appropriate measures have to be implemented
during the pre-installation to effectively reduce noise in the exam room.

2.7 Windows and curtains

When the examination room has a window with an aperture outside of the controlled light area
(day light, other...) a curtain has to maintain the light intensity under a limit fixed to 150 lux.
Chapter 3 Physical Requirements of Room
Direction 5314528-2-1EN, Revision 1