GE Innova 2100-IQ Preinstallation Manual page 31

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Installation safety instructions:
Storage safety instructions:
Chapter 2 Innova Systems
Contractor responsibility:
Electrical contractor is responsible for providing and connecting the
cables and configuring the PDB in by-pass mode.
GEHC is responsible for powering on the system with the UPS in
by-pass mode.
GEDE is responsible for UPS comissioning.
After removing the sidewalls of the UPS, make sure that all earth
connections when reassembling, are correctly reattached.
This UPS is intended for use in a controlled indoor environment free of
conductive contaminants and protected against animals intrusion.
HIGH GROUND LEAKAGE CURRENT: Ground connection is essential before
connecting to AC input!
For Europe only, if a differential breaker is placed on the hospital main
supply, upstream the PDB, the differential shall be set to 300 mA.
Switching OFF the unit does not isolate the UPS from the utility.
Do not install the UPS in an excessively humid environment or near water.
Avoid spilling liquids on or dropping any foreign object into the UPS.
The unit must be placed in a sufficiently ventilated area; the ambient
temperature should not exceed 104 F (40 C).
Optimal battery life time is obtained if the ambient temperature does not
exceed 77 F (25 C).
It is important that air can move freely around and through the unit. Do not
block the air vents.
Avoid locations in direct sunlight or near heat sources.
Check local regulations for UPS installation.
Store the UPS in a dry location; storage temperature must be within –13 F
(–25 C) to 131 F (55 C).
If the unit is stored for a period exceeding 3 months, the battery must be
recharged periodically (time depending on storage temperature).
Direction 5314528-2-1EN, Revision 1
31

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3100Innova 3100-iqInnova 4100Innova 4100-iq

Table of Contents