Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Table 3-8:
(cont'd)
Immunity Test
IEC 60601–1–2 Test
Electrical fast
2 kV for power supply
transient/burst
lines
IEC 61000–4–4
1 kV for input/output
lines
Surge
1 kV line(s) to lines(s)
IEC 61000–4–5
2 kV line(s) to earth
0 % U n for 5 sec
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000–4–11
3 A/m
Power frequency
(50/60 Hz) magnetic
field
IEC 61000–4–8
Note: These are guidelines. Actual conditions may vary.
Chapter 3 Physical Requirements of Room
Compliance Level
Level
2 kV for power supply
lines
1 kV for input/output
lines
1 kV line(s) to lines(s)
2 kV line(s) to earth
0 % U n for 5 sec
3 A/m
1A/m
Direction 5314528-2-1EN, Revision 1
Electromagnetic Environment
Mains power quality is that of a typical commercial and/or hospital
environment
Mains power quality is that of a typical commercial and/or hospital
environment.
Mains power quality is that of a typical commercial and/or hospital
environment. If the user of the Innova system requires continued
operation during power mains interruptions, it is recommended
that the Innova system be powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields are at levels characteristic of a
typical location in a typical commercial and/or hospital environment.
At that disturbance amplitude level the monitor image might present
some slight flicker. For image quality improvement, the monitor might
be removed from the low frequency magnetic field source vicinity.
At that disturbance amplitude level the system is fully operational.