Mounting Requirements; Floor Loading And Recommended Mounting Methods - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

2 Mounting Requirements

2.1 Floor Loading and Recommended Mounting Methods

See
Table 5-2
not specified in
in
Chapter 2, Section 3,
Table 5-2:
NET WEIGHT
PRODUCT OR
KG (LBS)
COMPONENT
Liberty LC
750 (1655)
Positioner
Table
590 (1300) See
note 1
900 (1983) See
note 1
IMPORTANT NOTICE
The Table baseplate or the System baseplate are mandatory to install the table (patient support). The Table must never be installed on grade.
299 (659)
C2 Cabinet
506 (1115.5)
C1 Cabinet
204 (450) See
COOLIX 4000
note 3
Detector Chiller
13.6 (30)
Thermo-Tek
Detector Chiller
14.6 ( 32.2)
Thermo-Con
Fluoro UPS UL
530 (1169)
(optional)
Fluoro UPS CE
480 (1059)
(optional)
124
. To obtain floor loading and recommended mounting methods for components
Table 5-2
, refer to the appropriate component Pre-Installation Manual listed
Basic Innova LC System Compatibility.
DIMENSIONS MM (INCHES)
WIDTH
DEPTH
See Illustrations INNOVA LC
POSITIONER DIMENSIONS:
Side view
Top view
Front view
in
Section 1,
Dimension Drawings
See illustrations in
Section 1,
Drawings
Omega IV
Omega V
Tilt
See Illustration C2 Cabinet Dimensions
in
Section 1,
Dimension Drawings
See Illustration C1 Cabinet Dimensions
in
Section 1,
Dimension Drawings
556 (21.9)
830 (32.7)
See note 2/3
See note 3
See Illustration Detector Chiller
Thermo–Tek Dimensions in
Dimension Drawings
See Illustration Detector Chiller
Thermo–Con Dimensions in
Dimension Drawings
690 (27.0)
800 (31.5)
680 (26.7)
800 (31.5)
Direction 5314528-2-1EN, Revision 1
LOAD
BEARING
AREA MM
HEIGHT
(INCHES)
Circle
diameter 600
(23.62)
Dimension
571.5x429
(22.5x16.9)
750x600
(30x24)
600x900
(23.63x35.44)
600x900
(23.63x35.44)
1090 (43)
Four casters
See note 3
283x351
Section 1,
(11.14x13.82)
13.5x11.14
Section 1,
(344x283)
1820 (70.1)
1450 (57.1)
WEIGHT/
OCCUPIED
MOUNTING METHOD
AREA
Recommended:
Through-Bolts
(12)
Alternates:
On Grade 5/8
in, Anchors
(12)
Above
Grade 3/4
in. Anchors
(12)
See
Illustration 5-28
,
Illustration 5-29
Illustration 5-30
2410 kg/m 2
Same as Innova
(492.3 lb/ft 2 )
LC Positioner See
IMPORTANT NOTICE
below.
2000 kg/m 2
(409 lb/ft 2
478 kg/m 2 (98
lb/ft 2 )
838 kg/m 2
(172 lb/ft 2 )
Not applicable
975 kg/m 2
Ground -mounted
(200 lb/ft 2 )
883 kg/m 2
Ground -mounted
(181 lb/ft 2 )
2 Mounting Requirements
and

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3100Innova 3100-iqInnova 4100Innova 4100-iq