GE Innova 2100-IQ Preinstallation Manual page 132

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
The Table baseplate or the System baseplate are mandatory to install the table
(patient support).
The Table must never be installed on grade.
NOTE:
For more details on Floor plate or Baseplates, refer to
location in concrete floor in
NOTE:
With any kind of fixation methods (Bolts M20, Mechanical anchors or Chemical anchors),
the number of holes used mandatory is:
Gantry baseplate : 12 max and 8 min holes used are acceptable
Table baseplate : 10 max and 8 min holes used are acceptable
Floor baseplate : 24 max and 12 min holes used are acceptable
we can have only 2 consecutive holes omitted.
NOTE:
In case of floor plate configuration, the gantry baseplate shall be mounted onto the floor plate,
but the table baseplate shall not be mounted on the floor plate.
Pull out efforts and recommendations about chemical anchors not provided by GE.
The following table provides the recommended chemical anchors for Omega/LC baseplates
and for the floor plate ordered locally that they could be used instead of bolts provided by GE.
Table 5-4: Chemical anchors Pull out efforts and recommendations
Mark
I on
Pull out effort
736 daN per bolt if 12
used and 1992 daN per
Number of holes in the
12 max (8 min mandatory) 10 max (8 min mandatory)
plate
Recommended
Supplier HILTIHVU
chemical anchors
adhesive capsule +
example 1
HAS Anchor rod
Threaded rod
M16 A4-70 / 333 131 5/8
Hole diameter in the
floor
Hole depth in the floor
132
Chapter 3, Section 3,
Gantry baseplate
Table baseplate Omega
Illustration 5-32
J on
1120 daN per bolt if 10
used and 2000 daN per
bolt if 8 used
Supplier HILTIHVU
adhesive capsule + HAS
Anchor rod
M20 A4-70 / 333 135 3/4
18 mm
(11/16 in)
125 mm
(5 in)
Structural Requirements.
Illustration 5-32
A on
4432 daN per bolt with 4
bolt if 8 used
Supplier HILTIHVU
adhesive capsule + HAS
Anchor rod
M20 A4-70 / 333 135 3/4
24 mm
(7/8 in)
170 mm
(6-5/8 in)
Direction 5314528-2-1EN, Revision 1
Illustration 5-33
and illustration Holes
Table Omega
Floor plate (to be ordered
Illustration 5-33
K on
272 daN per bolt if 24 used
bolts
and 2008 daN per bolt if 12
4 mandatory
24 max (12 min mandatory)
Supplier HILTIHVU adhesive
capsule + HAS Anchor rod
M16 A4-70 / 333 131 5/8
24 mm
(7/8 in)
170 mm
(6-5/8 in)
2 Mounting Requirements
locally)
Illustration 5-32
used
18 mm
(11/16) in
125 mm
(5 in)