Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Illustration 5-15: Floor Space Diagram for Lytron Chiller
The maximum elevation difference between X–Ray Tube Housing and Chiller is
around 3 meters (10 ft).
Exact dimensions of the Coolix 4000 chiller depend on model type. Required floor space
depends on ambient room temperatures. When in doubt, allow for maximum floor space.
112
Direction 5314528-2-1EN, Revision 1
1 Dimension Drawings