GE Innova 2100-IQ Preinstallation Manual page 48

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Table 3-3:
Core System
Room
Exam
Innova LC positioner and
Room
table
DL user area with 1 TFT
monitor
Ctrl Room
2 B&W flat monitors
C1 Cabinet
C2 Cabinet
Coolix X-Ray tube chiller
Tech
(1) (2)
Room
Detector conditioner
Main disconnect panel -
PDB
Total for core system
Options (3 & 5)
Room
3 in room B&W TFT
monitors
Exam
In room AW TFT monitor
Room
Typical injector
AW work station
Ctrl Room
2 AW TFT monitors
Printer
Typical configuration without fluoro UPS
(1) Air flow requirements 1200 m 3 /h (706 CFM)
NOTE:
NOTE:
(2) For more details, consult appropriate pre–installation manual
NOTE:
(3) For UPS 20 kVA option refer to
NOTE:
(4) Moderate use: 8 cases / 10 hours, typical use: 11 cases / 10 hours, maximum use:
during case.
NOTE:
(5) For IVUS enviromental limatations and heat output, refer to
48
Stand by
Moderate Use (4)
kW
BTU/hr
0.41
1394
0.55
0.16
546
0.16
0.17
573
0.17
0.41
1398
0.69
0.29
989
0.33
2.53
8619
4.49
0.21
709
0.21
0.4
1534
0.45
4.6
15762
7.0
Stand by
Moderate Use (4)
0.25
859
0.12
409
Same values as Stand
0.09
320
0.35
1201
0.24
818
0.31
1054
5.9
20424
Section 2.2.2
Direction 5314528-2-1EN, Revision 1
HEAT OUTPUT
Typical Use (4)
kW
BTU/hr
kW
1858
0.89
546
0.16
573
0.17
2366
0.99
1125
0.54
15309
5.49
709
0.21
1534
0.45
24020
8.9
Typical Use (4)
Same values as Stand
by
8.4
28682
10.3
2 Environmental Requirements/Limitations
Maximum Use (4)
BTU/hr
kW
3020
1.62
546
0.16
573
0.17
3389
1.29
1828
0.87
18725
6.93
709
0.21
1534
0.45
30324
11.7
Maximum Use (4)
Same values as Stand
by
by
34985
13.1
Section 2.2.3
BTU/hr
5517
546
573
4412
2966
23625
709
1534
39881
44543