Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
The PDB shall include the switch allowing the commutation of generator supply to the UPS.
NOTE:
Neutral shall be fed by the hospital. The customer provides the cables A and B.
Refer to the UPS vendor Service manual Installation Guide SG Serie 10, 20, 30 & 40 kVA for
NOTE:
more details
Chapter 6 Electrical Connections
Direction 5314528-2-1EN, Revision 1