2285053 - Through Bolts With Insert - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
2.3.3.5 2285053 – Through bolts with insert
Table 5-12:
Item
Name
Bolt, Hex
Washer, Flat
Plate
Nut, Hex
Grout assy
Dowel
Water
Electric
Separator
144
Part #
2296892
Through Bolt; M20 - 500-400
12 bolts for LC Positioner and 4 bolts
for Table
99142204
Special Flat Washer for Through
Bolts; one for each bolt
2290941
Special Steel Spacer Plate; 4 in. x
4 in. (102 mm x 102 mm); one for
each bolt
99141607
Hex Nut M20 STL galvanized two for
each bolt
2285055
- 10 kg Powdered Mortar Ardex K-15
- RTV Silicon Rubber Adhesive
- 18 mm making tape adhesive
- Open cell foam
2290937
Wood Dowel; 24 mm diameter
2268647
Manufactured part introduced in
Positioner key hole to fit 1 electrical
conduit of 6".
Direction 5314528-2-1EN, Revision 1
Description
16 pc
16 pc
16 pc
32 pc
1 kit
12 pc
1 pc
Quan.
Notes
Through Bolt mounting
method hardware
same as above
same as above
same as above
Used in constructing LC
Positioner grout dam
Electrical and water conduits
separated inch size.