Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
5.2 IVUS Floor Conduits
A 3 inches and max. 25 m long conduit between Control and Patient Rooms shall be
prepared below the floor for the IVUS Innova interconnection cables.
•
Locate the position of the IVUS components in the Control Room, according to the
workflow requirement of the operators.
•
Establish a 76 mm (3") conduit below floor from the IVUS CPU location in the Control
Room, to the foot end area of the Patient Table in the Procedures Room:
•
•
•
Use similar solutions (i.e. protection against cleaning fluids, etc. ) as it is used for
the Mac-Lab conduit
•
See also guidance on Typical Layout drawing 4-58f sheets E1 and S2, showing
Mac-Lab and P/M Gas location recommendations.
Chapter 6 Electrical Connections
Locate the Procedures Room exit of the conduit close to the centerline of the
Patient Table
Locate it as close to the Patient Table base as possible, considering :
•
Room for access to the table for service purposes
•
Limitation from other cables connected to the table
•
Location of the table plate
•
Location of Mac-Lab conduit
•
Location of Physio or Med Gases connections
•
The IVUS cables are moving cables, they have to allow the Patient Table
motions (Panning, Lifting and Rotating)
Direction 5314528-2-1EN, Revision 1