Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Access to a working phone at the facility
for emergency use, including MR magnet
7
delivery.
All walls primed (final coat not needed
on Day 1), and counter tops that will
8
support equipment must be installed.
No dust-producing cabinetry work in
installation areas.
Mechanical supplier has been provided
with a set of equipment installation
drawings for reference. For California,
9
permitted construction drawings or
PMI-specified installation drawings are
required.
Conduit/electrical cable ducting/dividers/
access flooring installed, with the
exception of surface-mounted floor
ducting. Wiring to the main disconnect
10
panel is installed and compliant with
equipment installation drawings or
pre-installation manual.
Issued Date: 7/9/07 Rev 11
214
X
X
X
X
X
X
X
X
GEHC Only: COE # (888) 799.7266 Option 5 (PMI Support)
Direction 5314528-2-1EN, Revision 1
X
X
5 Pre-Installation Checklist