Physical Run - System Core Detail - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
16.3 PHYSICAL RUN – SYSTEM CORE DETAIL
Table 6-4:
Cable
MIS
Assembly
number
Group n 1 (From C1 / C2 cabinets to LC positioner / Table)
27663A
2285897
27664A
2380494
27665A
2331836
27666A
2352342
27667A
2252342
27468A
2258191
27473A
2258191
27670A
2231647
27510A
2261135
27511A
2261135
27512A
2337267
27513A
2250247
27514A
5139188
27675A
2361531
27567A
2337165
27660A
2258294
27661A
2258294
27662A
2350071
27669A
2266532
27671A
2337252
27673A
2337252
27672A
2337252
27674A
2266584
Following cables are part of component
27586A
2260420
2363958
27722A
27589A
2258207
27590A
2115483
Group no.2 (From C1 / C2 cabinets to Control Booth)
27676A
2350102
2259592
27472A
27492A
2264834
27493A
2337270
192
Voltage rating
UL Style
(V)
2789
600
2463
600
2463
300
1019
600
1019
600
2789
300
2464
300
2789
30
Type CL2
150
2464
300
600
600
30
2464
300
2789
30
2789
30
2789
30
2464
300
300
Locaflex spec.
1311418A
Locaflex spec.
1311418A
300
4299
300
1354
Category 5
150
Cable
Max Voltage
diameter
carried (V)
(mm)
13.8
8.3
9.2
WATER HOSE
WATER HOSE
9.1
9.1
WATER HOSE
WATER HOSE
FIBER OPTIC
7
7.3
8.2
9.9
6.5
6.5
7
16.2
10.9
10.9
10.9
10.6
15
11.7
11.7
8.3
250
5.5
Direction 5314528-2-1EN, Revision 1
Connector type
DB 11 pin
Metrimate 6 pin
Metrimate 6 pin
Pre-stripping, ring terminal
Pre-stripping, ring terminal
HES 15 pin
CEI 320 - HBL4570C plug
DB 26 pin
Pre-stripping
DB 15 pin
Pre-stripping
Pre-stripping
AMP 8 pin RJ45
Circular connector
Amplimite 100 pin
Amplimite 100 pin
Amplimite 100 pin
DB 25 pin
AMP 25 pin
HV plug
HV plug
SCSI.50
8 plug power strip CEI320 -
(HBL4570C plug - ring terminal)
RJ45
16 Physical Runs
Bigger Plug
size (mm)
34.4
29
29
12
12
38.6
41.2
8.2
41.9
47.8
84.6
84.6
84.6
56.6
55
72
72
60
39
14.1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3100Innova 3100-iqInnova 4100Innova 4100-iq

Table of Contents