Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
8 Room Speaker
The remote speaker provides the operator in the exam room, the indication of record
exposure sequence (same level information as provided by the control room control panel).
The speaker must be installed in the exam room out of the patient vicinity to give easy access
to the FE for maintenance operations and to medical staff to adjust the tone signal level and
to be able to see the power LED indicator.
Illustration 6-15:
THE REMOTE SPEAKER KIT SHALL NOT BE INSTALLED IN THE PATIENT VICINITY.
Patient Vicinity Definition : see Patient Environment Equipment in
1,
Room Layout Considerations
174
Direction 5314528-2-1EN, Revision 1
Chapter 4, Section
8 Room Speaker