5. To proceed to the programming screen, press STEP FWD (B).
The programming screen will appear, with the file number and
step number at the top of the screen.
FGa
FILE# 1
1. INIT% =
2. FIN% =
6. To set the initial pressure or flow for this step, press INIT (1) to
activate this parameter. Use the number keys to enter the
desired value, then press ENTER to save it.
7. To set the final pressure or flow for this step, press FINAL (2) to
activate this parameter. User the number keys to enter the
desired value, then press ENTER to save it.
8. Set either the desired RATE (3) of change or DURATION (4) in
minutes. Once one value has been set and saved, the other will
automatically appear.
DURATION in minutes can have a resolution of 0.1, with a
maximum of 9,999 minutes per step.
9. If you want to add another step to the file program, press
INSERT (C).
10. The step number will increase by one, and the default initial
value will be the final value entered for the previous step. Edit as
desired.
11. When programming is complete, press the STORE key to save
the file and return to the home screen.
12. To start the program, press RUN two times.
When a gradient run is started, digital output '8' of the controller
ACCESSORY connector will toggle from high to low (open to
closed) for one second.
The flow rates and ramp rate for Pump A in each step will be in direct
opposite proportion to the values set for Pump B (INIT%B, FIN%B,
and RATE).
An entire gradient program can be removed only by deleting each of
its steps one at a time, as discussed in the next section, Section 4.5
"Review, Revise, & Hold Options". When the last remaining step is
deleted, the entire file is removed.