2
3
4
Figure 1-6 Pump rear panel connectors
Figure 1-6 #
Connector
1
Pressure outputs
2
Ground
3
Pressure transducer The pressure transducer cable must be plugged in for the pump to operate.
4
Control cable
5
6
External Fuses
7
Chassis ground
a. Internal fuses not replaceable by the operator: F101: 4.0 Amp "T", F102 & F104:1.5 Amp "T", and F103: 0.75 Amp "T".
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
7
1
Table 1-9 Pump Rear Panel Connectors
Two red binding post/banana jacks providing pressure output voltages.
Pump
65x
All others
A black binding post/banana jack providing a connection to circuit common.
This cable connects the pump to the controller.
Mains
IEC power connector with EMI filter.
a
Limits pump current drawn from main power supply.
Replace with same type: ("T" time delay fuses)
1) 2.0 Amp for 100/117 volt operation
2) 1.0 Amp for 234 volt operation
To remove, rotate cap counterclockwise.
Ground point for high static installations.
Section 1 Introduction
Description
Left Jack VDC/psi
Right Jack VDC/psi
1.0 V/2000 psi
1.0 V/1000 psi
6
5
0.1 V/2000 psi
0.1 V/1000 psi
1-13