Gradient Upload Commands - Two-Pump Flow Gradient; Upload Commands - Single-Pump Pressure Programming - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

Step Number 01
LGUL,F:xx,S:01,FL:xxx.xxxx,IB%:xx.xx,FB%:xx.xx,D:xxxx.x,
Step Numbers 02 through 99
LGUL,F:xx,S:xx,IB%:xx.xx,FB%:xx.xx,D:xxxx.x,
Figure 8-10 Gradient upload commands - two-pump flow gradient
Figure 8-11 Upload commands - single-pump pressure programming
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Gradient Total Flow Rate
Units = milliliters/minute
Range = (000.0000 – Maximum Flow Rate)
Step Number 01
File (01 – 99)
Final %B
Initial %B
Range = (00.00 – 100.0)
Step Number (02 – 99)
File (01 – 99)
LGUL,F:xx,S:xx,I:xxxxx,F:xxxxx,D:xxxx.x,
Step Number (01 – 99)
File (01 – 99)
Gradient Duration
Final %B
Range = (00.00 – 100.0)
Initial %B
Range = (00.00 – 100.0)
Gradient Duration
Range = (0000.1 – 9999.0 minutes)
Range = (00.00 – 100.0)
Pressure Duration
Final Pressure
Range = 00010 – Maximum
Units = psi
Initial Pressure
Range = 00010 – Maximum
Units = psi
Section 8 Serial Interface
Range = (0000.1 – 9999.0 minutes)
Range = 0000.1 – 9999.0 minutes
8-33