500Xv Technical Specifications - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

CYLINDER CAPACITY
POWER REQUIREMENTS
(Mains voltage line cord is a
"Disconnect Device")
LINE FREQUENCY
LINE VOLTAGE NOISE TOLERANCE
DIMENSIONS
b
WEIGHT
FLOW RATE RANGE
FLOW RATE ACCURACY
FLOW RATE DISPLAY RESOLUTION
ANALOG OUTPUT ACCURACY
DISPLACEMENT RESOLUTION
REFILL TIME
REFILL OR DEPRESSURIZATION RATE
PRESSURE RANGE
PRESSURE ACCURACY
PRESSURE REPEATABILITY
ZERO PRESSURE DRIFT
PRESSURE DISPLAY RESOLUTION
AMBIENT TEMPERATURE RANGE
TEMPERATURE DRIFT
HUMIDITY
DEAD (HEADSPACE) VOLUME
POLLUTION DEGREE
INSTALLATION CATEGORY
MAXIMUM ALTITUDE
a. Underwriters Laboratories (UL) has certified all SyriXus series syringe pumps with the exception of the 100 VAC versions.
b. A team lift is recommended when moving this instrument.
c.
Using water at 137.9 bar and a temperature controlled environment at 30 °C.
d. The analog output is an optional accessory.
e. Pressure repeatability specification is based upon re-zeroing pressure transducer every 48 hours. Refer to Section 3.8.11
"ZERO PRESS" in Section 3 "Basic Programming and Operation" for re-zeroing procedure.
f.
Volume in and above the piston seal, head clearance at automatic shutoff, and inlet and outlet ports to the fittings.
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Table 1-4 500xv Technical Specifications
507.38 mL
a
100 ± 10 VAC, 5A maximum
117 ± 12 VAC, 5A maximum
234 ± 23 VAC, 2.5A maximum
50 or 60 Hz
1.7
nominal rms line voltage, 10 µsecond pulses, any phase
angle, random or repetitive
PUMP
Width: 27.18 cm
Depth: 46.74 cm
Height: 103.00 cm
PUMP
44.6 kg
Refer to Figure 1-2.
c
± 0.5% (maximum 1.0 µL/min seal leakage)
1.0 µL/min
d
± 1% of selected range
25.36 nL /step
1.25 minutes
1.0 µL/min to 408 mL/min at any pressure from 0 to 344.7 bar
0.6895 to 344.7 bar
± 0.5% of full scale at constant temperature
e
± 0.5% of full scale within 48 hours at constant temperature
± 0.25% of full scale within 48 hours at constant temperature
6.895 kPa
5 to 40 °C
± 0.15% of full scale/°C
95% maximum
f
10.45 ± 0.65 mL
2
II
2000 m
Section 1 Introduction
}
Factory Set
CONTROLLER
Width: 27.18 cm
Depth: 30.48 cm
Height: 13.59 cm
CONTROLLER
2.96 kg
1-5