To Install the
Mounting Tabs
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 9 Pump Maintenance, Troubleshooting, and Servicing
5. The limit sensor assembly includes two limit sensors, the wire
harness and connector, and a conduit which protects the wires
running between the sensors as shown in Figure 9-17.
Depending upon when your pump was manufactured, the
protective conduit may be directly fastened to the pump or
retained with mounting tabs and cable ties.
If the conduit is fastened directly to the case, then it must be
removed and the plastic mounting tabs included with the limit
sensor assembly package should be installed using the existing
screws and screw holes.
a. Remove the screws and the conduit.
b. Then use the same screws and holes to install the mounts.
Be sure the tab portion of the mount is facing away from the
sensors, as shown in Figure 9-18, (the screw aperture is clos-
est to the sensors).
If your pump already has the plastic tabs installed, cut and
discard the plastic cable ties which run through the tabs and
hold the conduit.
6. Remove the four screws holding the limit sensors, and unplug
the cable from the power circuit board. The replacement limit
sensor assembly includes both the upper and lower sensor and
the plug.
There are several additional cable ties along the path from the limit
sensors to the power board. Two are reusable twist ties and (usually
three) others need to be cut to free the limit sensor wires. These
cable ties must be replaced to prevent the wires from entangling
with the worm gear.
7. Install the new limit sensor harness; and secure the protective
conduit by running the cable ties through the mounts and tight-
ening. Then cut the excess cable tie and discard.
Be sure the circuit board connector is mated pin-for-pin with the
jack.