Continuous Flow Air Valves Overview; Continuous Flow Technical Specifications-Air Valves - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

SyriXus Series Pumps Installation and Operation Guide
Section 5 Continuous Flow Introduction, Installation, and Operation
Table 5-2 Continuous Flow Technical Specifications-Air Valves
Pressure fluctuation at switchover
Minimum
Maximum system backpressure
Air supply source pressure
Maximum temperature
Maximum flow rate (mL/min)
Liquids: 65% of the single-pump
maximum rate
Wetted materials in valve package
Air valves
Tubing and fittings
5.2.1 Continuous Flow
Air Valves
Overview
EAR99 Technology Subject to Restrictions Contained on the Cover Page
5-2
5 psi (0.35 bar), at system backpressures from 100 psi (6.9 bar) to the
single-pump maximum. Higher fluctuation occurs at pressures below 100 psi.
50.76 psi (3.5 bar)
The single-pump maximum. Valves rated to 10,000 psi [689.5 bar]
(65x to 20,000 psi [1379 bar])
80 to 115 psi (5.5 to 7.9 bar)
Maximum system fluid temperature 320 °F (160 °C)
Maximum valve assembly temperature 150 °F (65 °C)
1000x
500xv
500x
260x
65x
Hastelloy, PEEK, and PTFE
Hastelloy, gold
The single air valves are designed to be installed on either side of a
single pump. A bracket at the bottom is inverted and the air supply
lines are rearranged to accomplish this. Instructions for mounting the
valve on either side are provided with the valve assembly, and are also
available at www.teledyneisco.com. The single valve assembly allows
the pump to continuously operate, but the fluid flow is interrupted
when the cylinder becomes empty and the pump refills. Delivery
resumes once the cylinder is refilled.
The dual air valve packages are installed between two pumps and
provide continuous flow with minimal interruption. When the deliv-
ering pump approaches empty, the other pump gradually takes over
fluid delivery allowing the first pump to refill. When the second pump
approaches empty, the first pump gradually takes over fluid delivery
allowing the second pump to refill. This cycle repeats continuously
until the fluid supply reservoir is empty or until the user (or external
programmed control) stops the pump.
Except for the 500xv, all SyriXus series pumps are available with
either single or dual air valves. Because of its specific purpose, the
500xv is only available with the dual air valve package.
265.2
132.6
132.6
69.55
19.50

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Syrixus 260xSyrixus 500xSyrixus 500xvSyrixus 1000x

Table of Contents