SyriXus Series Pumps Installation and Operation Guide
Section 3 Basic Programming and Operation

3.6 Menu Three

Figure 3-7 Menu 3 program selections

3.6.1 Poor Fill Alarm

3.6.2 Diagnostic Menu

3.6.3 Pressure
Calibration
Figure 3-8 Location of transducer midpoint value
EAR99 Technology Subject to Restrictions Contained on the Cover Page
3-10
Menu 3 provides programming options for poor fills, diagnostics,
pressure calibration, alternative pressure inputs, pressure control
setting, and pressure setpoint tolerance.
To save all settings and return to the main menu, press
.
/
(D)
PREVIOUS
RETURN
1. POOR FILL ALARM
2. DIAGNOSTIC MENU
3. PRESS. CALIBRATION
In Constant Pressure Mode, this feature allows you to set a fill point
as a percentage of pump volume. If this volume percentage is not
reached after a refill and re-pressurization, the system issues an
alarm and stops the pump.
This selection displays an additional menu containing testing options
for the system. Refer to Section 9.6 "Diagnostic Menu" for a complete
explanation of how to use the diagnostic menu.
This is a stored psi value to ensure optimal operation of the pressure
transducer. On most pumps, this value, known as the "midpoint
adjustment," can be found on a tag on the transducer cable (Figure
3-8).
The midpoint adjustment value is individual to each pump, and is
stored within the controller software. Therefore, this value must be
re-entered if the controller or pump is switched, controller is reset, or
if a pump is plugged into a different port on the controller.
Standard calibration for the transducer is a two-point adjustment, at
zero and maximum pressure. The midpoint adjustment number is the
difference between the pressure at half maximum, as read by a
pressure gauge with a precision of 0.25% accuracy, and the syringe
pump transducer output.
Press C
A, C
B, C
AL
AL
AL
the calibration number, and press
4. DIFF. MODES
5. PRESS. INTEGRAL
6. PRESS. DEADBAND
PREVIOUS
NOTE
C, or C
D to select the correct pump, enter
AL
.
ENTER