SyriXus Series Pumps Installation and Operation Guide
Section 3 Basic Programming and Operation
To Set Auto Refill
Volume
To Set Refill Rate
Disable/Enable Auto
Refill
EAR99 Technology Subject to Restrictions Contained on the Cover Page
3-6
If more than one pump is connected to the controller, pump B, pump
C, and pump D will be displayed on the fourth line, depending on
how many pumps are attached. To select a pump, press the softkey
under the pump designation.
1. Press ' 1' to set the volume for pump A (or the currently selected
pump). Refer to Section 3.4.2 "Pump Selection (if Multiple
Pumps)".
The units to the right of the symbol will blink, indicating that
you should enter a volume. Use the number keys to enter an
appropriate value and then press
1. Press ' 3' to set the refill rate for the designated pump. The refill
rate can also be changed from the main screen while the pump is
refilling.
2. A message will blink on the screen prompting to enter the
selected refill rate.
3. Enter the desired rate with the number keys and press
4. To save and exit the refill menu, press D, PREVIOUS.
The first line will display 'OFF' or 'ON', indicating whether or not this
feature is enabled for pump A (or the currently selected pump). Press
' 4' to toggle this feature OFF or ON for each pump.
Auto refill can be set independently for each pump.
If AUTO REFILL is ON, the pump will automatically switch to Refill
Mode when the volume reaches the auto refill mark. After refilling,
pumping will resume in the programmed mode. The ACCESSORY
outputs, which drive powered valves, will switch in sequence.
Unless otherwise specified, the pump will refill to full cylinder
capacity. To specify a smaller refill volume, press '2' and enter the
desired volume. Press '5' to toggle this feature OFF or ON for each
pump.
The system can also be programmed to refill a pump based on an
external analog input voltage, with a range of 0 to 11.5 volts. Infor-
mation about this feature is provided in Section 3.12.3 "External
Control for Refill".
NOTE
.
ENTER
NOTE
NOTE
.
ENTER