Teledyne SyriXus 65x Installation And Operation Manual page 101

Syringe pumps
Table of Contents

Advertisement

5.2.2 Air Valve
Installation
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 5 Continuous Flow Introduction, Installation, and Operation
An installation instruction sheet is provided with each valve package
with specific details related to the valve assembly. The various valve
packages are installed similarly, with slight differences in tubing con-
nections. The 500x / 260x valve package also includes steps to properly
attach the mounting bracket.
RISK OF INJURY. THE PRESSURES PRODUCED
COULD BE MORE THAN 1400 BAR. PLEASE UTILIZE
APPROPRIATE TUBING AND CONNECTIONS NOTED
IN THIS MANUAL.
To install the air valve package, refer to the installation guide sup-
plied with the valve assembly. A copy of this guide can also be down-
loaded from the Teledyne ISCO website. Figure 5-1 represents the
SyriXus 500x pump valve assembly; the other SyriXus series pump
models will be similar.
1. Select the proper pump tubing for the pumps being attached to.
These pump tubing assemblies are labeled and have the fittings
and ferrules already attached.
2. Place the pumps in the position where they are to be used. The
pumps are awkward to move once the valves have been installed.
3. Loosely attach the tubing lengths between the valve ports and
the pumps.
4. While supporting the valve assembly, align the screw holes in
the mounting bracket and loosely insert the screws and washers
provided (six places).
5. Adjust the position of the valve assembly and then tighten the
screws and fittings.
6. Trim the reagent supply tubing to the desired length, and then
attach the tubing to the check valve.
7. Attach the air supply line to the bottom of the actuator. The sup-
ply air pressure should be 80 to 115 psi [550 to 790 kPa].
8. Connect the actuator wires to the controller using Table
5-3 "ALL Dual Air SyriXus Pumps - Conversion Table for A-B
Pump Pair and C-D Pump Pair".
DANGER