5.3.2 Electric Valve
Installation
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 5 Continuous Flow Introduction, Installation, and Operation
An installation instruction sheet is provided with each valve package
with specific details related to the valve assembly. The valve package
installation is very similar between the pump models, with slight dif-
ferences in tubing connections.
The electric valve package is supplied with the mounting bracket and
pump tubing unattached. Parts for the other models are included in
the package but are not used. These parts should be kept so that the
valve can be mounted on other pump models should the need arise.
RISK OF INJURY. THE PRESSURES PRODUCED
COULD BE MORE THAN 700 BAR. PLEASE UTILIZE
APPROPRIATE TUBING AND CONNECTIONS NOTED
IN THIS MANUAL.
To install the electric valve package, refer to the installation guide
supplied with the valve assembly. A copy of this guide can also be
downloaded from www.teledyneisco.com. Figure 5-3 represents the
elevated temperature SyriXus 260x pump valve assembly; the other
SyriXus series pump models will be similar.
1. Select the proper pump tubing for the pumps being attached to.
These pump tubing assemblies are labeled and have the fittings
and ferrules already attached.
2. Attach the mounting bracket as appropriate for the pump model
being used, following the installation guide in the pump package.
3. Place the pumps in the position where they are to be used. The
pumps are awkward to move once the valves have been installed.
4. Loosely attach the tubing lengths between the valve ports and
the pumps.
5. While supporting the valve assembly, align the screw holes in
the mounting bracket and loosely insert the screws and washers
provided (six places).
6. Adjust the position of the valve assembly and then tighten the
screws and fittings.
7. Trim the reagent supply tubing to the desired length, and then
attach the tubing to the check valve.
8. Connect the cable below the motor housing on the back of the
valve assembly to the controller ELECTRIC VALVE connector.
Refer to Figure 1-6.