Command (Refer to
NOTE 1 & NOTE 2
.)
DIGITAL = xxxxxxxx
1
8
DIG CONTROL
DIG CONTROL = command
1
8
DISPENSEA
DISPENSEB
DISPENSEC
DISPENSED
DISPENSEA=#
DISPENSEB=#
DISPENSEC=#
DISPENSED=#
FLOW
FLOWCD
FLOWA
FLOWB
FLOWC
FLOWD
FLOW=#
FLOWB=#
FLOWC=#
FLOWD=#
G
GG
G&
G&2
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Table 8-3 Serial Commands (Continued)
Sets the digital output either High or Low, where 'x' is either 'H' or 'L'. The order cor-
responds with the outputs 1-8.
Returns the status of the digital output control bits as either REMOTE (R) or INTER-
NAL (I). The return message format is DIG CONTROL=xxxxxx, where 'x' is either
'R' or 'I'. "R" indicates the corresponding bit is controlled remotely; and 'I' indicates
the corresponding bit is controlled internally by pump software. The status order
returned corresponds with the outputs 1-8.
Sets the digital output control bits to either internal or remote, where 'x' is either 'R'
for REMOTE or 'I' for INTERNAL CONTROL. 'R' indicates the corresponding bit will
be controlled remotely (through the serial port). 'I' indicates the corresponding bit
will be controlled internally by pump software. The order corresponds with the out-
puts 1-8.
Returns the dispense volume for Dispense Mode.
Sets the dispense volume for Dispense Mode. Format is XXX.XXX mL. Leading
and trailing zeros are not required. Can only be changed when pump is stopped.
Returns the delivering pump's flow rate in Continuous Pumping mode and Modifier
Addition Mode. In INDEPENDENT Mode FLOW returns the pump A flow rate and
FLOWCD returns the pump C flow rate.
Returns the actual flow rate of the pump.
Returns the actual flow rate of the pump Enter # for a flow rate setpoint (Constant
Flow Mode). Format is XXX.XXXXXXX mL/min. Only 5 figures are significant. Lead-
ing and trailing zeros are not required.
Gets pump information. 'G' and 'GG' commands return a text string that contains
current pressure, analog input, and digital input information. 'G&' is the Get All com-
mand. This returns the same information as 'G,' plus flow rates, units, operation sta-
tus, and more. For four pump operation use the 'G&2' command. Refer to Section
8.8.1 "Get Status Command" for a complete description of this serial command.
Section 8 Serial Interface
Description
8-21