SyriXus Series Pumps Installation and Operation Guide
Section 2 Fluid System Connections and Accessories
FROM REFILL
KIT ASSEMBLY
spacer block
2-way
valve
TO
APPARATUS
Figure 2-7 Outlet valve package connection
EAR99 Technology Subject to Restrictions Contained on the Cover Page
2-10
When nuts are torqued to the cylinder cap ports, the pressure read-
ing may be affected. If the pressure no longer reads zero, release
the pressure in the cylinder and press ZERO PRESS. Refer to Sec-
tion 3.8.11 "ZERO PRESS" for instructions.
3. Use the valve fittings to attach stainless steel tubing to the top
port of the two-way valve.
a. For the SyriXus 260x, connect the reducing union to the
other end of this tubing.
4. Connect the stainless steel tubing between the valve's bottom
port and your apparatus. Cut to the desired length.
RISK OF INJURY. THIS EQUIPMENT PRODUCES
HAZARDOUS
APPROPRIATE TUBING AND CONNECTIONS AS
NOTED IN THE MANUAL.
Table 2-3 Manual Outlet Valve Kits
Pump Model
pre-bent
SST tubing
SST tubing
NOTE
DANGER
PRESSURES.
1000x
500x
260x
65x
PLEASE
UTILIZE
Part Number
68-1247-118
68-1247-082
68-1247-078
68-1247-126