SyriXus Series Pumps Installation and Operation Guide
Section 5 Continuous Flow Introduction, Installation, and Operation
5.5.5 To Run or Stop
Figure 5-9 Keystrokes to Run or Stop the pumps
EAR99 Technology Subject to Restrictions Contained on the Cover Page
5-18
2. T
P
– changes the time delay between the end of the
IME
OINT
pump stroke and the opening of the inlet valve. The default is 5
seconds but can be set to any value between 1 and 99 seconds.
3. F
NCV – toggles the No Check Valve option ON or OFF.
UNCTION
To enter the Pressure Point or Time Point:
1. Press the '1' key (for Pressure Point) or '2' key (Time Point) to
enter the settings mode for that option.
2. Using the keypad, enter the desired value between 010 and 250
for the pressure point, or between 1 and 99 for the time point.
3. Press the 'E
' key to store your selection.
NTER
4. Press the '3' key to toggle the NCV option ON or OFF. The last
value set is the mode that will be used.
5. Use the 'D' softkey (P
screen and exit the menu function.
With the NCV feature turned ON, there should be a noticeable delay
between the end of the pump delivery and the opening of the inlet
valve.
This menu setting allows the operator to run or stop both pairs of
pumps simultaneously.
(
MORE
(
MULTI
(F
OUR
R
/S
UN
Before pressing "run", ensure that "on cont flow" is displayed on the
screen, and that the set flow rate/pressure is correct.
) repeatedly to return to the main
REVIOUS
(P
REVIOUS
Press 3x to return
to run screen
)
-P
)
UMP
P
UMP
)
TOP
NOTE
)