Returned Values; Related Topics - Agilent Technologies 8960 Series Reference Manual

10 wireless communications test set
Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

830
ENTER A;Integrity,Channel_power
840
DISP "Test complete"
850
!
860
!***Display measurement results
870
PRINT "Integrity indicator = ";Integrity
880
PRINT "Minimum power = ";Channel_power;" dBm/1.23 MHz"
890
!
900
!***Return to default call processing states***
910
OUTPUT A;"CALL:CLPCONTROL:REVERSE:MODE ACTIVE" !Returns to active power control bits
920
OUTPUT A;"CALL:CONNECTED:DROP:TIMER:STATE 1" !Returns call drop timer to ON
930
OUTPUT A;"CALL:REGISTER:TIMER:STATE ON" !Returns timer-based registration to ON
940
OUTPUT A;"CALL:POW -50" ! Returns cell power to -50 dBm/1.23 MHz
950
!
960
DISP "Test complete"
970
OUTPUT A;"CALL:END" !Ends the call
980
END
990 Timeout: SUB Timeout
1000
PRINT "Program timed out"
1010
CLEAR 7
1020
CLEAR 714
1030
SUBEND

Returned Values

This query returns an integrity indicator followed by a channel power measurement result (if results are available).

Related Topics

"Channel Power Measurement Description" on page 44
"Channel Power Measurement Troubleshooting" on page 706
"CALibration Subsystem" on page 259
! Global timeout handler
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_prog_cpow.fm
Programming a Channel Power Measurement
157

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents