ABORt
ABORt[:ALL]
Function
Stops all measurements that are active.
If the trigger arm is set to single, the measurements will go to the idle state.
If the trigger arm is set to continuous, the measurements will re-arm and initiate again.
Setting
Range
• AFANalyzer - AF Analyzer measurement
• ATXPower - Analog Transmitter Power measurement
• CAPPower - Access Probe Power measurement
• CCTPhase - Code Channel Time/Phase error measurement
• CPOWer - Channel Power measurement
• CFERror - Frame Error Rate measurement
• CTXSpurious - TX Spruious Emmissions measurement
• FSTability - Frequency Stability measurement
• FM - Frequency Modulation measurement
• GPOWer - Gated Power measurement
• HWQuality - Handoff Waveform Quality measurement
• SAUDio - Swept Audio measurement
• TFERror - TDSO Frame Error Rate measurement
• WQUality - Waveform Quality measurement
Programming Example
OUTPUT 714;"ABORT:ALL" !Aborts all active measurements in progress.