SETup:TFERror
SETup:TFERror:CONFidence:REQuirement[:RATio]
Function
Sets/queries the test data service option frame error rate (TDSO FER) confidence requirement.
Setting
Range: 0.10 to 5.00 percent
Resolution: 0.01 percent
Query
Range: 0.10 to 5.00 percent
Resolution: 0.01 percent
*RST Setting
1.00
Programming Example
OUTPUT 714;"SETUP:TFERROR:CONFIDENCE:REQUIREMENT:RATIO 0.50
SETup:TFERror:CONTinuous
Function
Sets/queries the trigger arm state for the test data service option frame error rate (TDSO FER)
measurement. The trigger arm state determines whether the TDSO FER measurement will be
made once then stop (0|OFF (single)), or automatically re-arm upon completion of one
measurement and repeat the process (1|ON (continuous)).
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Programming Example
OUTPUT 714;"SETUP:TFERROR:CONTINUOUS OFF
!Sets measurement to single trigger mode.
SETup:TFERror:COUNt
Function
Sets/queries the number of frames to be counted during a test data service option frame error
rate (TDSO FER) measurement.
Setting
Range: 512 to 999,936
Resolution: integer multiples of 512
Query
Range: 512 to 999,936
Resolution: integer multiples of 512
*RST Setting
512
Programming Example
OUTPUT 714;"SETUP:TFERROR:COUNT 1536"
566
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_tferror.fm
"
"