Agilent Technologies 8960 Series Reference Manual page 642

10 wireless communications test set
Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

SYSTem:CORRection
SYSTem:CORRection:GAIN
Function
Sets/queries the up to 20 comma separated offset gain values in units of dB. Each gain value
corresponds with a frequency setting. See
See
"Measurement Related Configuration" on page
The units, dB, are optional. If no units are specified, units default to dB.
Sending a null list (no values) sets the state of all offsets to Off.
Range: 0 to 20 values between − 100 to +100
Setting
Resolution: 0.01
Range: 1 to 20 values between − 100 to +100, and NAN (9.91E+37)
Query
Resolution: 0.01
Factory
Offsets 1 through 20, are set to 0.00 dB, state is set to off
Setting
Programming Example
OUTPUT 714;"SYSTEM:CORRECTION:GAIN -3.12,-3.68,-4.23"
!Sets up to 20 comma separated offset values.
OUTPUT 714;"SYSTEM:CORRECTION:GAIN"
!Sets the state for all frequencies and offsets to Off.
SYSTem:CORRection:POINts
Function
Query returns the number of offsets that are in their On state when the RF IN/OUT Amplitude
Offset State is also set to On.
See
"Measurement Related Configuration" on page
Query
Range: 1 to 20, and NAN (9.91E+37)
Resolution: 1
Programming Example
OUTPUT 714;"SYSTEM:CORRECTION:POINTS?"
!Returns the number of frequencies and offsets with their state set to On.
642
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_system_correction.fm
"SYSTem:CORRection:FREQuency"