RFANalyzer:MANual:POWer[:SELected]
Function
Sets/queries the receiver expected input power when the test set's receiver input power control
mode is set to manual. See
The optional [:SELected] keyword in this command specifies that the expected input power
setting being queried applies to the current system type.
The test set's receiver power control mode is set by the
command.
The test set's measuring receiver will consider any amplitude offsets that are defined along with
this setting to calculate input power range settings.
Setting
Range: Test application and system type dependent (approximately -25 to +43 dBm).
Resolution: 0.01 dBm
Query
Range: Test application and system type dependent (approximately -25 to +43 dBm).
*RST setting
-75 dBm
Programming Example
OUTPUT 714;"RFANALYZER:MANUAL:POWER:SELECTED 24 DBM" !Sets the manual power control mode
expected input power level for the currently selected system type to 24 dBm.
RFANalyzer:MANual:ANALog:POWer[:SELected]
Function
Sets/queries the expected power when the mobile is on an analog voice channel. The selected
band is currently limited to the cellular band (800 MHz).
Setting
Range: -60 dBm to +53 dBm (interacts with the Amplitude Offset setting to determine the
available range)
Resolution: 0.01 dB
Query
Range: -60 dBm to +53 dBm
Resolution: 0.01 dBm
*RST setting
13 dBm
Programming Example
OUTPUT 714;"RFANalyzer:MANual:ANALog:POWer:SELECTED 24 DBM"
"Digital receiver power"
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_rfanalyzer.fm
.
"RFANalyzer:CONTrol:POWer:AUTO"
RFANalyzer
499