Agilent Technologies 8960 Series Reference Manual page 288

10 wireless communications test set
Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

CALL:D2KTest
CALL:D2KTest:ESNumber:HEX
Function
Sets/queries the test mode ESN (Electronic Serial Number).
Setting
Range: Up to eight characters containing alphanumerics 0123456789ABCDEF or NULL
Query
Range: String of up to eight characters containing alphanumerics 0123456789ABCDEF or NULL
*RST Setting
"00000000"
Programming Example
OUTPUT 714;"CALL:D2KTEST:ESNUMBER:HEX 'ABCD1234'" !Sets the mobile station's ESN to ABCD1234.
CALL:D2KTest:QPCHannel:INDicator[:PAGing]
Function
Sets/queries the QPCH (Quick Paging Channel) paging indicators. Setting this parameter to
ALL1 will cause the slotted mode mobile station operating in the idle state to monitor the paging
channel during each assigned slot. When this parameter is set to ALL0 the mobile station will
not monitor the paging channel.
Note: QPCH configuration change indicators are not settable. The test set will always transmit
configuration change indicators as 0 bits.
Setting
Range: ALL0|ALL1
Query
Range: ALL0|ALL1
*RST Setting
ALL0
Programming Example
OUTPUT 714;"CALL:D2KTEST:QPCH:IND ALL0" !Sets the paging indicators to 0 (off).
288
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_d2ktest.fm