CALL:DCONnected:TIMeout
Function
Sets/queries the maximum time the test set will wait for a hanging CALL:DCONnected:STATe?
query to complete. Default setting units are seconds. To set timeout time in units other than
seconds, include an optional unit specifier in the command string.
A timeout timer is started whenever the call-state-change detector becomes armed or gets
rearmed when already armed. The duration of this timeout is set using the
CALL:DCONnected:TIMeout command and should be greater than the maximum amount of
time the control program needs to wait between arming the call-state-change detector and the
connect/disconnect operation starting. Once the process starts and the data channel's state has
moved to one of the transitory states, the protocol timers take over and prevent the data
channel's state from staying in a transitory state forever. See
Synchronization"
Setting
Range: 0 to 100 seconds
Resolution: 0.1 seconds
Query
Range: 0 to 100 seconds
Resolution: 0.1 seconds
*RST Setting
10 seconds
Programming Example
OUTPUT 714;"CALL:DCONNECTED:TIMEOUT 3"
!Sets the CALL:DCONnected:STATe? query timeout time
OUTPUT 714;"CALL:DCONNECTED:TIMEOUT 500 MS"
Sets the CALL:DCONnected:STATe? query timeout time
!Sets the
.
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_dconnected.fm
"Call Processing State
to 3 seconds.
to 500 milliseconds.