AFGenerator
AFGenerator:COUPling
Function
Sets the output of the audio generator to be ac or dc coupled to the front-panel AUDIO OUT port.
Setting
Range: AC or DC
Query
Range: AC|DC
*RST setting
DC
Programming Example
OUTPUT 714;"AFGENERATOR:COUPling AC" !Sets the audio generator coupling to AC.
AFGenerator:FREQuency
Function
Sets/queries the frequency of the audio generator. The units (HZ|KHZ|MHZ|GHZ) are optional.
If no units are specified, then units default to HZ.
Setting
Range: 1 Hz to 20 kHz
Resolution:0 .1 HZ
Query
Range: 1 Hz to 20 kHz
Resolution: 0.1 Hz
*RST setting
1 kHz
Programming Example
OUTPUT 714;"AFGENERATOR:FREQUENCY 1000" !Sets the audio generator frequency to 1000 Hz.
AFGenerator:PULSe[:STATe]
Function
Sets/queries the audio generator pulse state.
The pulse state must be on when the test set's audio generator is used for audio stimulation
during a decoded audio measurement.
When the state is on, the audio signal from the test set is pulsed at a 10 Hz rate with a 50% duty
cycle.
Setting
Range: 0 | OFF | 1 |ON
Query
Range: 0|1
*RST setting
0 (off)
Programming Example
OUTPUT 714;"AFGENERATOR:PULSE ON" !Sets the audio generator pulse to ON.