SETup:WQUality
SETup:WQUality:COUNt[:SNUMber]
Function
The setting form of this command performs two functions: (1) sets the number of waveform
quality measurements to be made in a multi-measurement, and (2) sets the multi-measurement
STATe command to 1 (on).
The query form returns the number of waveform quality multi-measurements to be made when
multi-measurement state is ON.
Setting
Range: 1 to 999
Query
Range: 1 to 999
*RST Setting
Off (10 when STATe = On)
Programming Example
OUTPUT 714;"SETUP:WQUALITY:COUNT:SNUMBER 5"
!Turns on multi-measurements and sets the number of measurements to 5.
SETup:WQUality:COUNt:NUMBer
Function
Sets/queries the number of waveform quality measurements to be made when the
multi-measurement state is ON.
Setting
Range: 1 to 999
Query
Range: 1 to 999
*RST Setting
10
Programming Example
OUTPUT 714;"SETUP:WQUALITY:COUNT:NUMBER 5"
!Sets the number of measurements to 5 when the multi-measurement state is set to ON.
SETup:WQUality:COUNt:STATe
Function
Sets/queries the multi-measurement state of the waveform quality measurement.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Programming Example
OUTPUT 714;"SETUP:WQUALITY:COUNT:STATE ON" !Turns on multi-measurements.