Agilent Technologies 8960 Series Reference Manual page 529

10 wireless communications test set
Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

SETup:CPOWer:CONTinuous
Function
Sets/queries the trigger arm state for the channel power measurement. The trigger arm state
determines whether the channel power measurement will be made once then stop (0|OFF
(single)), or automatically re-arm upon completion of one measurement and repeat the process
(1|ON (continuous)).
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Programming Example
OUTPUT 714;"SETUP:CPOWER:CONTINUOUS OFF
!Sets measurement to single trigger mode.
SETup:CPOWer:COUNt[:SNUMber]
Function
The setting form of this command performs two functions: (1) sets the number of channel power
measurements to be made in a multi-measurement, and (2) sets the multi-measurement STATe
command to 1 (on).
The query form returns the number of channel power multi-measurements to be made when
multi-measurement state is ON.
Setting
Range: 1 to 999
Query
Range: 1 to 999
*RST Setting
Off (10 when STATe = On)
Programming Example
OUTPUT 714;"SETUP:CPOWER:COUNT:SNUMBER 5"
!Turns on multi-measurements and sets the number of measurements to 5.
SETup:CPOWer:COUNt:NUMBer
Function
Sets/queries the number of channel power measurements to be made when the
multi-measurement state is ON.
Setting
Range: 1 to 999
Query
Range: 1 to 999
*RST Setting
10
Programming Example
OUTPUT 714;"SETUP:CPOWER:COUNT:NUMBER 5"
!Sets the number of channel power multi-measurements to 5.
"
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_cpower.fm
SETup:CPOWer
529