SETup:AFANalyzer:TIMeout[:STIMe]
Function
Sets/queries the AF analyzer measurement's timeout setting and automatically sets the timeout
state to ON.
Setting
Range: 1 to 999 seconds
Resolution: 1 second
Query
Range: 1 to 999
*RST Setting
10 seconds, on
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:TIMEOUT:STIME 5S"
!Sets timeout state to ON and sets the timeout time to 5 seconds.
SETup:AFANalyzer:TIMeout:STATe
Function
Sets/queries the AF analyzer measurement's timeout state.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:TIMEOUT:STATE ON"
SETup:AFANalyzer:TIMeout:TIME
Function
Sets/queries the AF analyzer measurement's timeout setting to be used when the timeout state is
set to ON.
Setting
Range: 1 to 999 seconds
Resolution: 1 second
Query
Range: 1 to 999
*RST Setting
10 seconds
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:TIMEOUT:TIME 5S"
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_afan.fm
SETup:AFANalyzer
511