Shutdown

Before disconnecting the ARCADIS Varic from the mains, you must shut it down.
During the shutdown procedure the imaging system is powered down. There are
two different procedures, each with a different reset time.
The shutdown procedure used depends on how you switch off the
ARCADIS Varic:

Simple shutdown

Press the OFF switch on the monitor trolley.
– The C-arm system is immediately switched off.
– The imaging system and monitor A switch off after the system has been
shut down.
The image system is shut down to "hibernate" before it is switched off by using
the OFF button. When the ARCADIS Varic is switched on again, it will normally
be ready for operation after approx. one minute.
Depending on the configuration, it may take longer than one minute for the
ARCADIS Varic to be ready for operation again.
It is possible to switch on again in approx. 5 seconds after you used the switch
off button (even during shut down).
Switching off (simple shut down) can be used up to 15 times in a row. After the
tenth shut off using the OFF button, a dialog opens in which you can select sim-
ple or complete shut down. After the 15th switch off using the OFF button,
a complete shut down automatically occurs with appropriate notification.
If you want to ensure a very short power-up time of your ARCADIS Varic system,
we recommend that you initiate a complete shutdown yourself before the maxi-
mum number of 15 simple shutdown procedures is reached. You can do this dur-
ing a break between examinations, for example, and then switch the
ARCADIS Varic on again in time before the next examination.