Checks Prior To Special Examinations (E.g. Of The Open Heart And Skull); Dose And Consistency Test - Siemens ARCADIS Varic Operator's Manual

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

ARCADIS Varic
SPR2-310.620.30.02.02
Checks prior to special examinations (e.g. of the open
heart and skull)
Make sure that there is an additional conductive connection between the
C-arm system and a point of potential equalization, e.g. the patient table.

Dose and consistency test

The ARCADIS Varic system features automatic object position detection.
Proceed as follows:
Place the object or filter into the beam path.
Release fluoroscopy in automatic mode for approx. 10 s.
Press the ADR stop button on the control panel.
Attach the dose measuring chamber to the image intensifier input.
Release fluoroscopy again to perform the dose measurement.
Maintenance
0
0
5
of 16

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents